58 . Lambda XLS/XLS+ User's Guide
9. Press Next .
The Standard Curve – Standards screen is displayed.
10. Enter the concentration values for each standard using the keypad numbers and the up
and down arrows to move between the different standard boxes.
The available range is from 0.001 to 9999.
11. Press Next
.
The Standard Curve – Calibration screen is displayed.
NOTE: If any duplicate or non-monotonic entries are entered on the Standards screen, when
Next is pressed the instrument will beep and highlight the incorrect entry.