Configuring Security Settings
Configuring WLAN Network Settings
3-12 Efficia CM Series Network Configuration Guide
• Server Address: Enter the IP address or host name of the destination server.
The following settings are optional:
•
Server Port: If necessary, change the default port number (4000) to the port number
used by the destination server.
• Export to cfg File: Select this check box if you use the Import Settings and Export
Settings features to import configuration settings on multiple monitors and if all
monitors export data to the same server port number. The server port number is
saved to the .cfg file. For more information about exporting and importing
configuration settings, see the Efficia CM Series Patient Monitors Installation and
Configuration Guide.
•
SSL: The Efficia CM Series monitor can use the secured socket (SSL/TLS) to make
secured connections to the server and to encrypt data. To use SSL/TLS, a root CA
certificate that can verify the server certificate must be installed before connection.
Measurement units:
• Blood Pressure Units
• Temperature Units
• CO2 Pressure Units
• CO2 (Gas)
• AGT/N20 (Gas)
• SpHb Units (Masimo only)
• Height Units
• Weight Units
Specify that the above mentioned measurement units in the exported data match
the corresponding units of measurement on the receiving system, regardless of the
setting on the monitor.
Note — For information on the Save Patient Data option, see the Efficia CM Series Patient
Monitors Installation and Configuration Guide.
To configure network settings for the destination server:
Step
1
On the Admin tab of the System Menu, select Data Export.
The Data Export Menu appears.
Note — The available options in the Data Export list vary. If your monitor is
wireless, the list contains a LAN/WLAN option; if your monitor is not wireless,
the list contains a LAN option.
2
Select LAN/WLAN from the Data Export list.
The Server Address, Server Port, Export to cfg File, and SSL options become
available.
3
Select Server Address.
A keyboard opens.