Regulatory and Safety Information
Efficia CM Series Network Configuration Guide 7-3
Radio Regulatory Compliance
2. Increase the separation between the affected equipment and monitor, or monitor
and receiver.
3. Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
4. Consult the IT department, service personnel, or biomedical engineer for help.
Caution FCC Caution: Any changes or modifications not expressly approved by the party
responsible for compliance could void the user’s authority to operate this equipment.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following
two conditions: (1) This device may not cause harmful interference, and (2) this device
must accept any interference received, including interference that may cause undesired
operation.
This device is restricted to indoor use when operated in the 5.15 to 5.25 GHz frequency
range.
FCC requires this product to be used indoors for the frequency range 5.15 to 5.25 GHz to
reduce the potential for harmful interference to co-channel Mobile Satellite systems.
This device does not permit operations on channels 116-128 (5580 – 5640 MHz) for 11na
and 120-128 (5600-5640 MHz) for 11a which overlap the 5600 -5650 MHz band.
Note — FCC Radiation Exposure Statement:
This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environment. This equipment should be installed and operated with
minimum distance 20 cm between the radiator and your body.
Industry Canada
Warning RF Radiation Hazard Warning
To ensure compliance with FCC and Industry Canada RF exposure requirements, this
device must be installed in a location where the antennas of the device will have a
minimum distance of at least 20 cm from all persons.
European Union
This device complies with the essential requirements of the Radio Equipment Directive
2014/53/EU. The following test methods have been applied in order to prove
presumption of conformity with the essential requirements of the Radio Equipment
Directive 2014/53/EU:
Interface Specifications
Frequency Bands 2.4 GHz:
ETSI: 2.4 to 2.483 GHz
FCC: 2.4 to 2.483 GHz
KC: 2.4 to 2.483 GHz
MIC: 2.4 to 2.495 GHz
5 GHz:
ETSI:
5.15 to 5.35 GHz
5.47 to 5.725 GHz
FCC:
5.15 to 5.35 GHz
5.47 to 5.725 GHz
5.725 to 5.85 GHz
KC:
5.15 to 5.35 GHz
5.47 to 5.725 GHz
5.725 to 5.825 GHz
MIC:
5.15 to 5.35 GHz
5.47 to 5.725
GHz
Maximum Output
Power for Each
Frequency
20.00 dBm, 2.4G: 2.412~2.472 GHz
22.20 dBm, 5.15-5.25 GHz
22.10 dBm, 5.47-5.725 GHz