Configuring WLAN Network Settings
Efficia CM Series Network Configuration Guide 3-19
Wireless Specifications
Guidance and Manufacturer’s Declaration
The Efficia CM patient monitors are intended for use in the electromagnetic environment
specified in the following tables. The customer or the user of the product should verify
that it is used in such an environment.
Table 3-1
Emissions Test Compliance Electromagnetic Environment Guidance
RF emissions
CISPR 11
Group 1 Efficia CM Patient Monitors may contain an
internal radio which emits RF energy. Including
these radio modules, total RF emissions are
very low and not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A The Efficia CM monitors are suitable for use in
a Hospital Environment, whose power is not
directly connected to the public low-voltage
power supply network.Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Table 3-2
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
+ 6 kV contact
+ 8 kV air
+ 6 kV contact
+ 8 kV air
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
+ 1 kV
differential
mode
+2 kV common
mode
+ 1 kV
differential
mode
+2 kV
common
mode