Wireless Specifications
Configuring WLAN Network Settings
3-20 Efficia CM Series Network Configuration Guide
Surge
IEC 61000-4-5
+ 1 kV
differential
mode
+ 2 kV common
mode
+ 1 kV
differential
mode
+ 2 kV
common
mode
Mains power quality should be
that of a typical hospital
environment.
Voltage dips,
short
interruptions,
and voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5% U
T
(> 95% dip in
U
T
)
for 0.5
cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(> 95% dip in
U
T
)
for 5 sec
< 5% U
T
(> 95% dip in
U
T
)
for 0.5
cycle
40% U
T
(60% dip in
U
T
)
for 5 cycles
70% U
T
(30% dip in
U
T
)
for 25 cycles
< 5% U
T
(> 95% dip in
U
T
)
for 5 sec
Power
frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
hospital environment.
Note: U
T
is the AC mains voltage prior to application of the test level.
Table 3-2
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
Guidance