Configuring WLAN Network Settings
Efficia CM Series Network Configuration Guide 3-21
Wireless Specifications
Table 3-3
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Efficia CM patient monitors, including
cables, than the recommended
separation distance calculated from
the equation applicable to the
frequency of the transmitter.
Conducted RF
IEC 61000-4-6
3 Vrms
0.15 to 80 MHz
3 V rms Recommended Separation Distance
Radiated RF
IEC 61000-4-3
3 V/m
80 to 2500 MHz
3 V/m
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol:
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mo-
bile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the Efficia CM patient monitor is used
exceeds the applicable RF compliance level above, the Efficia CM patient monitor should be observed to verify
normal operation. If abnormal performance is observed, additional measures are necessary, such as re-orienting
or relocating the monitor.
b
Signal levels above 1V/m in the 1215 – 1240 MHz band may disrupt wireless communications from the Efficia CM
Patient Monitor if the operating channel in the 2.4 GHz band is at twice the frequency.
d
3.5
3
------- P=
; 0.150 to 80 MHz
d
3.5
3
------- P=
; 80 to 800 MHz
d
7
3
--- P=
; 800 to 2500 MH