Serial Configuration Overview
Configuring Serial Network Settings
4-2 Efficia CM Series Network Configuration Guide
Serial Configuration Overview
To configure serial data export, follow these steps:
Verifying the RS-232 Serial Port Settings
The RS-232 serial port settings on the receiving device must match the settings on the
monitor. Verify that the receiving application is configured as follows:
Baud Rate: 38400 bps
Data Bits: 8
Stop Bits: 1
Parity: None
Flow Control: None
Enabling Serial Data Export
To enable serial data export:
Step See
1
Verify the serial port settings. “Verifying the RS-232 Serial Port
Settings” on page 4-2.
2
Enable serial data export. “Enabling Serial Data Export” on
page 4-2.
3
Connect the serial adapter. “Connecting the Serial Interface
Adapter” on page 4-3.
Step
1
Select the System button in the toolbar to open the System Menu.
2
Select the Admin tab.
3
In the window that appears, enter the Administrator password, 2-1-5 (or the
password you configured) and select OK twice.
The Admin tab appears.
Caution
The Admin tab remains unlocked for 1 minute after you close it. This
allows you to open the menu again without having to re-enter the
password. Do not leave the monitor unattended during the unlock time.
4
Select Data Export.
The Data Export Menu appears.
5
Select Serial from the Data Export list.