Time Synchronization
Efficia CM Series Network Configuration Guide 5-3
Enabling Time Synchronization
7 Select the time synchronization type for the monitor:
If the monitor synchronizes its time with the timestamp of an HL7
acknowledgment message from the receiving system, select Data
Export
. Proceed to step 10.
If the monitor synchronizes its time with an SNTP server, select SNTP.
SNTP Server Address and Time Zone become available.
8 Select SNTP Server Address.
A keyboard appears.
9 Enter the IP address or the hostname (up to 50 characters) of the SNTP
server, and then select
OK.
10 Select the
Time Zone field to display the list of time zones. Select the time
zone that matches your region.
11 Select Apply and close the Date/Time Settings Menu.