Managing Logs
Troubleshooting
3-42 Efficia CM Series Repair Guide
Managing Logs
The monitor collects data in the logs described in the following table.
Caution To avoid loss of data, export the log contents before the log becomes full.
Log Description
User Activity Log The following events are logged:
• Setting changes (except diagnostics)
• User actions in menus (except diagnostics)
• Night Mode entry or exit
• Waveform Freeze
• Recording events
• C.O. measurement
• IBP and CO
2
Zero
The log contains up to 3000 entries. The oldest entry is
deleted when full capacity is reached.
The exported log is named: UserActivity.csv.
Alarm Log The following events are logged:
• All high, medium, low priority patient physiological
alarms, occurred and released
• All high, medium, low priority technical alarms,
occurred and released
• Alarm Silence/Audio Pause/Audio Off
• Alarm Silence from Central
Note — When the monitor is in Demo mode, events are
indicated by [Demo] at the end of the entry.
The log contains up to 3000 entries. The oldest entry is
deleted when full capacity is reached.
The exported log is named AlarmActivity.csv.
Warning
The alarm logs are retrievable after a controlled
shutdown. Manually performing the shutdown
procedure or allowing the battery to completely
discharge are considered controlled shutdowns.
However, the loss of AC power without battery backup
will result in the loss of the alarm logs.