Removing the Front End Assembly
Repairing the Monitor
4-20 Efficia CM Series Repair Guide
Removing the Front End Assembly
To remove the plastic patient connector panel from the front end assembly or SpO
2
board, you must remove the front end assembly from the main board assembly.
Step
1
Remove the rear case. See “Removing the Rear Case” on page 4-10.
2
Remove the three Torx T10 screws (torque: 0.8 Nm) from the front end
board.
3
Lift the front end board straight up and away from the connectors on the
main board.
Screws
Connectors