EasyManua.ls Logo

Philips Efficia CM Series - ECG Output Performance Test; NBP Test

Philips Efficia CM Series
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ECG Output Performance Test
Performance Verification Testing
2-22 Efficia CM Series Repair Guide
To perform the Respiration performance test:
ECG Output Performance Test
The ECG Output performance test checks the performance of ECG synchronization
between the monitor and a defibrillator. This test needs to be performed only if this
feature is in use at your facility.
To perform this test, you need:
Defibrillator with ECG sync cable
Patient simulator
To test the ECG output:
NBP Test
These tests check the performance of the non-invasive blood pressure measurement:
NBP accuracy
NBP calibration procedure (if required)
Step
1
Change the patient simulator configuration to: Respiration rate 40 rpm.
2
Verify that the displayed respiration waveform and respiration rate (± 2
rpm) are the same as the values you selected on the patient monitor:
between 38 – 42 rpm.
3
If you do not get the expected results, see “ECG/Respiration
Measurement Problems” on page 3-11.
Note — Philips employees record this value as R:P or R:F.
Step
1
Connect the patient simulator to the ECG connector on the monitor and
the defibrillator to the ECG output on the monitor.
2
Set the patient simulator to the following configuration:
HR = 100 bpm
ECG sinus rhythm
3
Turn on the defibrillator in simulation mode.
4
Verify that the ECG signal is displayed.
5
If you do not get the expected results, see “ECG/Respiration Measurement
Problems” on page 3-11.
Note — Philips employees record this value as EO:P or EO:F.

Table of Contents

Other manuals for Philips Efficia CM Series

Related product manuals