Required Test Equipment
Performance Verification Testing
2-4 Efficia CM Series Repair Guide
Required Test Equipment
The following table lists the additional test equipment that you need to perform each
of the tests in this chapter.
To Perform This Test You Need This Test Equipment
“Visual Test” on page 2-10 None
“Power-On Self Test” on page 2-11 None
“Alarms Test” on page 2-11 NBP cuff and hose
“SpO
2
Test” on page 2-12 Adult SpO
2
sensor
SpO
2
simulator
“Sidestream CO2 Calibration Test”
on page 2-13
Electronic flowmeter, M1026-60144
Gas calibration equipment:
Cal 1 gas 15210-64010 (5% CO
2
/95%
N
2
)
Cal gas flow regulator M2267A
Cal tube 13907A
“Respironics CapnoTrak Sidestream
CO
2
Accuracy Test” on page 2-16
Water filter assembly.
Large airway adapter.
Gas calibration equipment:
Cal 1 gas 15210-64010 (5% CO
2
/95%
N
2
)
Cal gas flow regulator M2267A
Cal tube 13907A
“Respironics LoFlo Sidestream CO
2
Accuracy Test” on page 2-18
Respironics CO
2
LoFlo adult airway adapter
kit.
Gas calibration equipment:
Cal 1 gas 15210-64010 (5% CO
2
/95%
N
2
)
Cal gas flow regulator M2267A
Cal tube 13907A
“Mainstream CO2 Accuracy Test”
on page 2-20
• CO
2
sensor
Gas calibration equipment:
Cal 1 gas 15210-64010 (5% CO
2
/95%
N
2
)
Cal gas flow regulator M2267A
Cal tube 13907A
M1990A airway adapters or similar (2)
“ECG/Respiration Performance
Test” on page 2-21
Patient simulator
ECG patient cable