Clearing Patient Data
Troubleshooting
3-44 Efficia CM Series Repair Guide
To delete the logs:Clearing Patient Data
Clearing Patient Data
If you return a monitor for service, you must first clear the patient data.
Patient data and a summary are automatically stored in internal memory. A summary
report can be downloaded to a USB flash drive.
The Clear Data function resets the user settings to the factory defaults and clears the
following:
All patient data
All profiles and profile settings
All logs
All cloneable settings, except the AC line frequency
Monitor name
The following settings are not cleared:
Monitor serial number
Key code
LAN/WLAN MAC Address
Touchscreen, temperature, IBP, and C.O. calibration data
Language
NBP cycle count
3
Select the View Logs button.
The View Log Menu appears.
4
Select the Manage Logs tab.
5
Select the Save All button.
All logs are saved to the USB flash drive as a tar file named: SN_logs.tar,
where SN is the serial number of the monitor.
6
Close the menu.
1
Access the Admin tab. See “Accessing the Admin Functions” on page 3-5
of the Efficia CM Series Service Guide.
2
Select the View Logs button.
The View Log Menu appears.
3
Select the Manage Logs tab.
4
Select the Clear All button.
All logs are deleted from the monitor.
5
Close the menu.