Removing the Philips SpO
2
Board
Repairing the Monitor
4-22 Efficia CM Series Repair Guide
Removing the Philips SpO
2
Board
Note — This procedure may vary slightly due to different monitor configurations.
To remove the SpO
2
board:
Step
1
Remove the rear case. See “Removing the Rear Case” on page 4-10.
2
Remove the front end assembly. See “Removing the Front End Assembly”
on page 4-20.
3
Remove the two white plastic #1 Phillips screws (torque: 0.1 Nm).