Performance Verification Testing
Efficia CM Series Repair Guide 2-11
Power-On Self Test
Power-On Self Test
To perform the power-on self test:
Alarms Test
The following test allows you to verify that the visual and audio alarms are working.
To perform this test, you need an NBP cuff and hose.
Step
1
Connect the monitor to an AC power source.
2
Press the On/Standby key to power up the monitor.
3
Make sure that the monitor powers up successfully as described in the
following sequence:
The color bar is displayed and then the LCD turns off.
The Charging LED turns on (when the battery is available).
A tone sounds and the Philips splash screen appears.
The Alarm LED lights in blue, followed by yellow, and then by red.
The Date/Time Menu appears to allow the configuring of date and time.
After you acknowledge the Date/Time Menu, the main screen
appears.The expected result is that the monitor starts up and displays
the main screen.
Note — If configured, the Admit Patient Menu appears. After you
acknowledge the Admit Patient Menu, the main screen appears.
The expected result is that the monitor starts up and displays the main
screen. For detailed information about the start-up and power
sequences, see “Start-up and Power Sequences” on page 3-2
If the LEDs do not function as expected, see “Power Problems” on
page 3-5.
If the display does not function as expected, see “Power Problems” on
page 3-5 or “Display Problems” on page 3-6.
If you do not hear a startup tone, or the monitor displays the Speaker
Malfunc
error message, see “Error Codes” on page 3-18.
Note — Philips employees record this value as PO:P or PO:F.