Performance Verification Testing
Efficia CM Series Repair Guide 2-9
Enabling Demo Mode
Enabling Demo Mode
Warning Do not connect a patient to a monitor running in Demo mode. Values represented in
Demo mode do not represent measurements from a patient connected to the
monitor, and may lead to incorrect diagnoses.
Demo mode is used to demonstrate the monitor without connecting the monitor to an
actual person. Demo mode simulates all patient parameters and generates alarms
when alarm settings are exceeded.
Caution Entering Demo mode clears the patient data.
To put the monitor in Demo mode:
4
In the window that appears, select the password field.
An onscreen keyboard opens.
5
Enter the password, 129, and select OK.
6
Select OK.
The Diagnostics Menu displays additional options.
Step
Step
1
Access the Admin tab. See “Accessing the Admin Functions” on page 2-6.
2
Select Demo Mode.
Note — Checking Demo Rack allows you to demonstrate the rack
measurements when Demo Mode is enabled.
3
Close the menu.
A confirmation window appears.
4
Select OK.
The monitor enters Demo mode and clears all patient data. DEMO
appears on the top of the monitor display.
5
To exit Demo mode, press the On/Standby key to turn the monitor off.
The monitor clears all simulated patient data.