Performance Verification Testing
Efficia CM Series Repair Guide 2-1
2
Performance Verification Testing
Overview
This chapter includes the performance tests that must be performed after repairs and
before the monitor is put back into service.
Notes
•
The procedures in this chapter assume knowledge of basic monitor
operation. For details about using the monitor, see the Efficia CM Series
Patient Monitors Instructions for Use.
• For more information about routine performance tests, see the Efficia CM
Series Patient Monitors Service Guide.
Testing and Inspection Guidelines
The following table lists the tests that Philips requires that you complete after
performing monitor repairs or software upgrades.
For information about routine maintenance procedures, see the Efficia CM Series
Patient Monitors Service Guide.
After Complete These Tests
Upgrading the software • Power-on
• Verify that your system settings
are preserved
Opening the monitor for any reason • Power-on
• Alarms Test
• Pneumatic leakage
• Speaker
• All safety tests
Replacing any internal parts (except NBP
parts, SpO
2
board, Sidestream CO
2
module)
• Power-on
• Pneumatic leakage
• Speaker
• All safety tests