EasyManua.ls Logo

Philips Efficia CM Series - Replacing a Fuse; Replacing the Recorder

Philips Efficia CM Series
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Replacing a Fuse
Repairing the Monitor
4-8 Efficia CM Series Repair Guide
Replacing a Fuse
To replace a fuse in the AC power connector:
Replacing the Recorder
To replace the recorder:
Step
1
Shut down the monitor. See “Shutting Down the Monitor” on page 4-1.
2
Remove the AC power cord from the rear of the monitor.
3
Remove the battery from the monitor (see “Replacing the Battery” on
page 4-5).
4
Use needle-nose pliers to pull the fuse holder out of the AC power
connector.
5
Replace the fuse or fuses, and then push the fuse holder back into the
power connector until it snaps into place.
Step
1
Shut down the monitor. See “Shutting Down the Monitor” on page 4-1.
2
Remove the AC power cord from the rear of the monitor.
3
Remove the battery from the monitor (see “Replacing the Battery” on
page 4-5).
4
Open the recorder door on the rear case assembly and remove the
paper roll.