Replacing the Power Entry Module
Repairing the Monitor
4-12 Efficia CM Series Repair Guide
Replacing the Power Entry Module
To replace the power entry module:
Step
1
Remove the rear case. See “Removing the Rear Case” on page 4-10.
2
Remove the two Torx T10 screws (torque: 0.8 Nm) from the power input
module.
3
Pull the module out of the chassis and disconnect the three cables from the
module.
Screws
Dangerous Voltage
symbol
Cables