Accessing the Diagnostics Menu and Maintenance Options
Performance Verification Testing
2-8 Efficia CM Series Repair Guide
Accessing the Diagnostics Menu and Maintenance Options
Some of the verification tests require you to use the Diagnostics Menu or the
Maintenance options.
Note — The Diagnostics Menu and Maintenance options are password protected.
After you access the Diagnostics Menu, you must enter an additional password to
access the Maintenance options. Only trained biomedical engineers or support
personnel should access the Maintenance options and perform maintenance and repair
procedures.
Caution When you open the Diagnostics Menu, monitoring is suspended and measurements
stop.
Notes
The Charging LED does not show battery status while the Diagnostics Menu
is open.
The Diagnostics Menu is not available when the monitor is in Demo mode.
Use the Diagnostics Menu to perform the system diagnostic and verification tests and
track component usage hours.
The Diagnostics Menu allows you to:
Configure the hardware and software;
View monitor and rack status and errors;
Perform measurement and monitoring functions;
View connected plug-ins’ status and errors;
Enter the system serial number if the monitor, rack, or plug-in module boards
have been replaced.
Note — You cannot access the Diagnostics Menu when the monitor is in Demo mode.
To access the Diagnostics Menu:
Step
1
Access the Admin tab. See “Accessing the Admin Functions” on page 2-6.
2
Select the Diagnostics button.
The Diagnostics Menu menu appears.
3
Select the Maintenance >> button to access additional options.

Table of Contents