Replacing the Battery
Repairing the Monitor
4-6 Efficia CM Series Repair Guide
Caution Store batteries at a room temperature of 25˚ C (77˚ F) or lower. Exposure to higher
temperatures may reduce battery life, damage the batteries, and degrade overall
monitor performance.
To replace the battery:
Step
1
Shut down the monitor. See “Shutting Down the Monitor” on page 4-1.
2
Remove the AC power cord from the back of the monitor.
3
Insert a slotted screwdriver into the slot in the battery cover.
4
Hold one hand over the battery cover while you twist the screwdriver
slightly to release the battery cover from the case.
Warning
Twist the screwdriver slowly and gently to avoid expelling the battery
cover with enough force to cause personal injury.
Slot