Troubleshooting
Efficia CM Series Repair Guide 3-3
Start-up and Power Sequences
User Action Expected Result
Connect the AC power cord or
insert a battery or batteries (or
both) into an unpowered
monitor.
Press the On/Standby key.
The color bar is displayed and then the LCD turns
off.
The Charging LED turns on (when a battery is
present).
A tone sounds and the Philips splash screen
appears.
The Alarm LED lights in blue, followed by yellow,
and then by red.
The
Date/Time Menu appears to allow the
configuring of date and time.
After you acknowledge the Date/Time Menu, the
main screen appears.
Press the On/Standby key
when the monitor (with AC
power and/or battery power) is
off.
The color bar is displayed and then the LCD turns
off.
The Charging LED turns on (when a battery is
installed).
A tone sounds and the Philips splash screen
appears.
The Alarm LED lights in blue, followed by yellow,
and then by red.
If configured, the
Admit Patient Menu appears.
After you acknowledge the Admit Patient Menu,
the main screen appears.
Software reset (system or user). The color bar is displayed and then the LCD turns
off.
The Charging LED turns on (when a battery is
installed).
A tone sounds and the Philips splash screen
appears.
The Alarm LED lights in blue, followed by yellow,
and then by red.
The main screen appears.
Plug AC power cord into a
monitor that is off (battery
only).
Press the On/Standby key.
The color bar is displayed and then the LCD turns
off.
The Charging LED turns on (when a battery is
installed).
A tone sounds and the Philips splash screen
appears.
The Alarm LED lights in blue, followed by yellow,
and then by red.
If configured, the Admit Patient Menu appears.
After you acknowledge the Admit Patient Menu,
the main screen appears.