Immersing Transducers
1 Immerse this section only, up to 5.1 cm (2 in) from the connector’s strain relief.
2 Maximum allowable immersion depth; you are not required to immerse to this depth if it is unnecessary.
Disinfectants Compatibility
Read this information before performing disinfection and sterilization procedures. It discusses
recommended disinfectants and choosing an appropriate disinfectant for the required level of
disinfection. You must check “Disinfectants and Cleaning Solutions Compatibility Table” on page
278 for the chemical compatibility of disinfectants and cleaners with specific transducers. In
addition, the table indicates if a device can be sprayed or wiped only, or if it can be soaked.
Transducer Care
Disinfectants Compatibility
276 EPIQ 7 User Manual 4535 617 25341