Inspect the transducer cable, case, and lens before each use. Check for cracks or other damage
that jeopardizes the integrity of the transducer. Report any damage to your Philips customer
service representative, and discontinue use of the transducer.
For all information on transducer cleaning and disinfection, see the “Transducer Care” section.
For all information about the use of acoustic coupling gels, see “Ultrasound Transmission Gels”
on page 184.
CAUTION
Some ultrasound coupling gels, as well as some solutions for precleaning, disinfecting, and
sterilizing can damage a transducer. Before using a gel or solution on a transducer, see
“Ultrasound Transmission Gels” on page 184 or the “Transducer Care” section. You can also
contact your local Philips representative. For contact information, see “Customer Service” on
page 22.
If you encounter poor image quality or transducer problems, see “Troubleshooting” on page
295.
Printer and DVD Recorder Maintenance
Before performing any maintenance on a device, observe the following warnings and cautions:
WARNING
If the device is internal to the system, turn off the system and disconnect the system from the
wall outlet. If the device is external to the system, disconnect the device from the wall outlet.
CAUTION
Do not scratch the roller or allow dirt and dust to contact the roller of a printer.
CAUTION
Do not use strong solvents such as thinner or acetone, or abrasive cleaners, because those
will damage a device cabinet.
System Maintenance
Printer and DVD Recorder Maintenance
294 EPIQ 7 User Manual 4535 617 25341