Cable Type Length Philips Part Number
Adult ECG leadset (IEC) – – 453561490141/M1672A
Pediatric ECG leadset (AAMI) – – 453561490101/M1624A
Pediatric ECG leadset (IEC) – – 453561490121/M1669A
S -video S-video cable Any –
LAN Twisted pair Any –
USB Shielded Any –
Approved Transducers for Electromagnetic Compliance
The imaging transducers used with the system may affect its emissions. The transducers listed
in “Clinical Applications and Transducers” on page 174, when used with the system, have been
tested to comply with the Group 1, Class B emissions, as required by International Standard
CISPR 11. Use only those transducers.
WARNING
Using cables, transducers, and accessories other than those specified for use with the system
may result in increased emissions from, or decreased immunity of, the system.
NOTE
The use of an Intracardiac Echocardiography (ICE) Probe or an Transesophageal
Echocardiogram (TEE) probe results in Class A Radiated Emissions and should be used in a
hospitable environment only.
Approved Accessories for Electromagnetic Compliance
Accessories used with the system may affect its emissions. The accessories listed here, when
used with the system, have been tested to comply with the Group 1, Class B emissions as
required by International Standard CISPR 11. Use only the accessories listed here.
Electromagnetic Compatibility
Safety
EPIQ 7 User Manual 4535 617 25341 73