When connecting other accessories to the system, such as a remote video monitor or
computer, it is the user’s responsibility to ensure the electromagnetic compatibility of the
system. Use only CISPR 11 or CISPR 22, Class B-compliant devices, unless otherwise noted.
WARNING
Using cables, transducers, and accessories other than those specified for use with the system
may result in increased emissions from, or decreased immunity of, the system.
CAUTION
Use only Class-B USB storage devices with the system. Some plastic-cased unshielded USB
devices may cause RF emissions that exceed Class-B limits. See the device’s documentation to
determine whether it is Class-B compliant.
Approved Accessories
Accessory Manufacturer Model Number
Ultrasonic imaging transducer Philips Use only the transducers listed in
“Clinical Applications and
Transducers” on page 174
DVD recorder Sony DVO-1000MD
Black-and-white printer Sony UP-D897MD
Color printer Sony UP-D25MD
Electromagnetic Immunity
The system is intended for use in the electromagnetic environment specified here. The
customer or the user of the system should ensure that it is used in such an environment.
NOTE
The guidelines specified here may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Safety
Electromagnetic Compatibility
74 EPIQ 7 User Manual 4535 617 25341