7. Yearly checkup and safety test
7.1 Safety inspection.
To prevent your Guidance from having hidden defects acquired during its lifecycle it is strongly
recommended to have the device checked regularly for its proper functioning by qualified personnel.
In some countries this is mandatory or your insurer may be requiring it. A record of these inspections
should be filed so the measurements can be compared with earlier values to alert for a possible
potentially dangerous tendency. To this purpose you can e.g. make copies of these pages.
/ This document is also available in the Guidance E/C guideline.
7.2 Inspection interval.
The design of the device is based on a yearly inspection. If the legislation of your country or your
insurer call for a shorter period you should adhere to the latter.
7.3 Inspection results.
The inspection is passed when all items as presented on the standard document on the next page are
passed. In the unlikely event the device fails the deviation should be fixed before recommisioning it.
Repairs are only to be made by technicians authorised by GymnaUniphy nv. Contact your supplier
on this matter.
7.4 Different current wave forms for checkup.
The following figures are the references for the different current wave forms as explained in the
checkup document.
Figure 1 Figure 2
Phyaction Guidance E/C – service manual version 0.5 Page 25