BT37 MarkII: Service Manual MA200606 1.0.0
Calibration using automatic calibration kit
16. Cal Report. Scroll Down to check Validity Field. If anything is invalid,
investigate the reason for the calibration failure and once any issues have been
corrected, start the calibration again. If OK, Apply adjustments. See adjusted
Offsets.
(Invalid)