Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-20 Planmeca Proline CC panoramic x-ray
Technical Manual
Start the Dimax3 calibration program by double-clicking the Dimax2Tool.exe program icon.
The Dimax2Tool window appears.
Figure 28
Calibrating the sensor head
Select Proline from the Settings / Type pulldown menu.
Figure 29
The calibration must be performed with 2x2 (high resolution), 3x3 (enhanced resolution) and 4x4
(normal resolution) binning. Select the binning from the Setting / Binning pulldown menu.
Figure 30
Click the Dimax3 panoramic exposure button. A dark current exposure (i.e. an exposure without
radiation) must be taken first and a message window shown below appears.
Figure 31
Dimax3 panoramic