Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-30 Planmeca Proline CC panoramic x-ray
Technical Manual
If the distances from the middle ball to the center of the tenth ball on both the right and left are
more than 122mm then the patient positioning mechanism is too far backward and must be
moved forwards (
Fig. 54).
Figure 54
If the distances from the middle ball to the center of the tenth ball on both the right and left are
less than 118mm then the patient positioning mechanism is too far forward and must be moved
backwards (
Fig. 55).
Figure 55