11.2 Warning message
Warning message Reason
Safety interlock open The absorber temperature has exceeded 100°C. In order to protect
the device, the safety Interlock is switched to the impermissible
operating status.
Tab. 11.2: Warning message
This warning message does not imply an error. The message is automatically
reset as soon as the absorber temperature is below 100°C again. The safety
interlock is then also automatically switched back to the permissible operating
status.
11.3 Charge status of the lithium-ion battery
The charge status of the battery is displayed in percent. The accuracy of this display
depends on various factors (e.g. temperature, age of the battery, etc.). Recharge the
battery when the charge level reaches 20percent.
The battery can be charged with a max. charging current of 1.3A. The battery has a
capacity of 1000mAh. The charging time of the battery on the PC via USB connection
with a charging current of 500mA is approx. 2hours.
At 100% charge status, the possible operating time is approx. 6hours (equivalent to
approx. 100measurements). When using all power saving functions (see chapter 14.4
on page37) approx. 15hours.

Table of Contents

Other manuals for Primes CUBE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Primes CUBE and is the answer not in the manual?

Primes CUBE Specifications

General IconGeneral
BrandPrimes
ModelCUBE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals