16 Maintenance and service
The operator is responsible for determining the maintenance intervals for the measuring
device.
PRIMES recommends a maintenance interval of 12months for inspection and calibra-
tion.
If the device is used only sporadically, the maintenance interval can be extended up to
24months.
16.1 Exchanging the protective window on the device
The protective window in the beam entrance is a wearing part and can be replaced
if necessary. Low levels of contamination of the protective window can be removed
carefully with Isopropanol when cooled down (observe the manufacturer‘s safety
instructions). In case of heavy, non-removable contamination or damage, the protective
window must be replaced with a new one.
The protective window is coated with an anti-reflective coating and has low
reflectance values of less than 1%. To avoid increased reflectance values, use
only original PRIMES protective windows.
Protective window diameter 55mm
Glass thickness 1,5mm
Order number 410-070-021 (1piece); 410-011-018 (10pieces)

Table of Contents

Other manuals for Primes CUBE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Primes CUBE and is the answer not in the manual?

Primes CUBE Specifications

General IconGeneral
BrandPrimes
ModelCUBE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals