35,0(6
Cube
29
Revision 10 EN - 03/2021
Absorber temperature in °C Min. energy input in J Max. energy input in J
20 50 3000
25 50 2800
30 50 2600
35 50 2400
40 60 2200
45 90 2100
50 140 1900
55 200 1700
60 290 1500
65 390 1300
70 520 1100
Tab. 12.1: Absorber temperature with the recommended minimum and permissible maximum
energy input (ambient temperature 20°C)
12.3 Waiting times until the next measurement in a measurement series
Please note that at high measuring frequencies the measuring accuracy may be limited.
For serial measurements within the specified accuracy, we recommend the following
waiting times until the next measurement.
Energy input in J Maximum repetition rate in s
50 40
100 50
200 60
1000 240
Tab. 12.2: Waiting times until the next measurement in series measurements