35,0(6
Cube
54
Revision 10 EN - 03/2021
21.3 Max. laser power depending on the irradiation time
for devices with advanced absorber
Maximum permissible irradiation time in s
Laser power in W
0 0.20.1 0.4 0.5 0.7 0.90.3 0.6 0.8 1.0 1.1
0
2000
4000
6000
8000
10000
12000
14000
21.4 Max. laser power depending on the beam diameter
for devices with advanced absorber
Beam diameter in mm
Maximum permissible laser power in W
100
1000
10000
0 5 15 3010 20 25 35
100
1000
10000
0 5 10 15 20 25 30 35
Empfohlene Werte
Grenzwerte
Recommended values
Limit values