14.4 Settings for trigger level, power saving functions and automatic
measurement readiness
The trigger level determines the sensitivity of the photodiode for laser detection. The
trigger level Low is recommended for pulsed beams.
With a low trigger level a measurement may be triggered even without a laser beam. In
this case set the trigger level to a higher value. The default setting for the Cube is trigger
level Low.
1. Set the Trigger level (Laser
power).
In the Cube the default setting
is the Trigger level Low (20W-
200W).
2. Click on the Save settings
button.
In the LaserDiagnosticsSoftware LDS
you can make additional settings for
power saving functions and the auto-
matic measurement readiness:
1. Enter the desired settings in the
Device control.
2. Click on the Save settings
button.
Autom. ready for measurement By default, the device automatically returns to
measurement readiness after each measurement. If
you uncheck the box, you will need to reset the unit
after each measurement by briefly pressing the on/of
button.

Table of Contents

Other manuals for Primes CUBE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Primes CUBE and is the answer not in the manual?

Primes CUBE Specifications

General IconGeneral
BrandPrimes
ModelCUBE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals