Owner’s Manual
AutoCode
®
Blood Glucose Meter AutoCode
®
Blood Glucose Meter
51
Do not get any liquids, dirt, dust, blood, control solu-
tion, or cleaning materials inside the meter through the
test strip port or the data port. Never spray cleaning
solution on the meter or immerse it in any liquid.
Lancing Device
Wipe your Prodigy lancing device
with a PDI Super Sani-Cloth Germi-
cidal Disposable Wipe. Do not im-
merse the lancing device in any liq-
uid. Do not use bleach. Disinfect your
Prodigy lancing device twice a week.
After cleaning the device, unscrew the cap; wipe it with
a PDI Super Sani-Cloth Germicidal Disposable Wipe for
two (2) minutes. Make sure the cap is completely dry
before reattaching it to the device.
Disinfecting your System
Disinfection is a process that destroys pathogens, such
as viruses and other microorganisms on the meter and
lancing device surfaces.
use bleach, harsh abrasives, or cleaning materials other
than the PDI Sani Cloths. The meter has been tested to
withstand cleaning for up to two times per week for 3
years (520 validated cycles (1 cycle = 1 cleaning wipe
and 1 disinfection wipe)).

Table of Contents

Other manuals for Prodigy Autocode

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Prodigy Autocode and is the answer not in the manual?

Prodigy Autocode Specifications

General IconGeneral
Measurement Range20-600 mg/dL
Sample Size0.6 µL
Test Time6 seconds
CodingNo Coding Required
Battery Type2 x AAA batteries
Automatic Shutoff2 minutes
Data PortUSB
Operating Conditions50-104°F (10-40°C)
Memory CapacityUp to 450 test results