Owner’s Manual
AutoCode
®
Blood Glucose Meter AutoCode
®
Blood Glucose Meter
57
Message Description
All error messages are listed.
Error
What it means: Appears when the
temperature is out of the system
operating range (50°F–104°F or
10°C–40°C).
Action: Repeat the test after the meter
and test strip are within the operating
temperature range.
What it means: Insufcient blood sample.
Test strip may be absorbing blood too
slowly.
Action: Insert a new test strip. If the
problem persists, please contact
Customer Care at 1.800.243.2636.