Owner’s Manual
AutoCode
®
Blood Glucose Meter
60
vThe control solution test is out of range.
Probable Cause Actions
Error in performing Read the instructions thoroughly
the test. and repeat the test again.
Control solution bottle Shake the control solution
not shaken well.
bottle vigorously and repeat
the
test again.
Expired or Check the expiration date and
contaminated the discard date of the
control solution. control solution.
Control solution that is Control solution should come
too warm or too cold. to room temperature before
testing.
Test strip Repeat the test with a new
deterioration. test strip.