Installation and setup
Page C-44 Operator’s Manual PiCCO
2
Version 3.1
6.1 Alarms
Alarms are displayed with priority 1, error messages with lower priority are overridden.
PCCO / PCCI > alarm limit
PCCO / PCCI < alarm limit
Battery status < 25% and
device is not connected to
mains power
Connect device to mains power
Internal error -
restart the device or
SERVICE
No signals longer than 10 s
(e.g. internal USB does not
work)
Restart the device, if error exists
further on, SERVICE necessary
Internal error -
SERVICE necessary
6.2 Error Messages
Text error messages are displayed in the information bar (chapter C4.1). When viewing old measurements the
respective error messages are displayed as well.
General
Enter patient data for
indexed values
Patient data (height / weight) etc. not
entered
Thermodilution
Invalid / Faulty catheter - Change
catheter
Faulty catheter, not
compatible or not connected
Connect compatible
catheter
Unknown catheter - Enter ACC
Unknown catheter type with
unknown resistor
Enter ACC in
corresponding field
Injection Error - Repeat injection,
rapid and smooth
Repeat injection rapidly
and smoothly
Injectate Temp. Sensor Error -
Check sensor cable and sensor
housing
Defective injectate temp.
sensor or sensor cable
Check and replace if
necessary