10.3 Cleaning
When cleaning the PiCCO
2
, please proceed as follows:
x Clean the surface of the PiCCO
2
with a soft, lint-free cloth moistened with warm water (40°C/104°F
maximum) and soap, a diluted non-caustic detergent, tenside, ammonia- or alcohol-based cleaning
agent.
x Do not use strong solvents.
x Take special care when cleaning the touch screen of the PiCCO
2
as it is more sensitive to rough
cleaning methods than the housing.
x Do not allow water or cleaning solution to enter the connectors of the monitor.
Recommended cleaning agents:
x Tensides Edisonite Schnellreiniger
®
x Ammonias Dilution of Ammonia < 3%, Window Cleaner
x Alcohol Ethanol 70%, Isopropanol 70%
10.4 Disinfection
Disinfect the product as stated by your hospital´s policy in order to avoid long-term damages.
Before disinfection clean the equipment.
CAUTION: Do not mix disinfecting solutions (such as bleach and ammonia) as hazardous gases
may be the result.
Recommended disinfecting agents:
x Alcohol based: Ethanol 70%, Isopropanol 70%, Cutasept
®
, Kodan
®
, Sagrosept
®
, Sterilium Fluid
®
x Aldehyde: Cidex
®