EasyManua.ls Logo

PULSION PiCCO2 - Battery Function and Status

PULSION PiCCO2
105 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation and setup
Operator´s Manual PiCCO
2
Page
Version 3.1 C-51
9 Battery Function
The device is equipped with a battery to ensure the maintenance of operations for at least 30 minutes without
mains power supply.
The PiCCO
2
must be connected to the mains power supply for at least 6 hours in order to completely charge
the battery.
In order to maintain the charging condition of the battery, the device must remain on mains power supply
during storage and the main switch at the back of the device must remain on position “ON”.
The battery symbol in the information bar shows the current state of charge:
Battery charging
Grey, last bar of status display is blinking
Fully charged, device connected to
mains power
No battery symbol
Battery-powered, > 25% capacity
Grey, remaining capacity in 25% steps
Battery-powered, < 25% capacity
Red, blinking with acoustic alarm
No battery in device
Grey, crossed out symbol
In cases of a battery alarm the acoustic alarm can be paused for 5 minutes.
CAUTION: Check the battery state of charge when using the device. Only when the battery is fully
charged, the use of the device can be ensured for the stated time without being connected to the
mains.

Table of Contents