EasyManua.ls Logo

PULSION PiCCO2 - Page 94

PULSION PiCCO2
105 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
Page E-10 Operator’s Manual PiCCO
2
Version 3.1
Table 6 Recommended separation distances between portable and mobile RF
communications equipment and the PiCCO
2
(EN 60601-1-2:2007; 5.2.2.2)
Recommended separation distances between
portable and mobile RF communications equipment and the PiCCO
2
The PiCCO
2
is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the PiCCO
2
can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the PiCCO
2
as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1.17 1/V *
P
80 MHz to 800 MHz
d = 1.17m/V *
P
800 MHz to 2.5 GHz
d = 2.33m/V *
P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection of structures, objects and people.
CAUTION: The PiCCO
2
is subject to specific precautions concerning EMC (Electromagnetic
compatibility) and is only allowed to be installed and used according the EMC advice contained
in this user´s manual.
Portable and mobile high frequency communication devices may influence the PiCCO
2
.
When high frequency devices are used during surgery, the applying standards for high
frequency devices for surgery have to be followed.

Table of Contents