Remote Control Commands
R&S
®
NRPxxS(N)
129User Manual 1177.5079.02 ─ 10
SYSTem:TRANsaction:END............................................................................................139
SYSTem[:SENSor]:NAME...............................................................................................139
SYSTem:VERSion?........................................................................................................140
SYSTem:COMMunicate:NETWork:RESTart
Effective only for the R&S NRP LAN power sensors.
Restarts the network connection to the DUT, that means terminates the connection and
sets it up again.
Example:
SYST:COMM:NETW:REST
Usage: Event
SYSTem:COMMunicate:NETWork:RESet
Effective only for the R&S NRP LAN power sensors.
Resets the LAN network settings to the default values.
Usage:
Event
SYSTem:COMMunicate:NETWork:STATus?
Effective only for the R&S NRP LAN power sensors.
Queries the network configuration state.
Example:
SYST:COMM:NETW:STAT?
Query
UP
Response: The network is active.
Usage: Query only
SYSTem:COMMunicate:NETWork[:COMMon]:DOMain <domain>
Effective only for the R&S NRP LAN power sensors.
Sets the domain of the network.
Parameters:
<domain>
Example:
SYST:COMM:NETW:COMM:DOM 'ABC.DE'
Sets ABC.DE as domain of the network.
SYSTem:COMMunicate:NETWork[:COMMon]:HOSTname <hostname>
Effective only for the R&S NRP LAN power sensors.
Sets the individual hostname of the sensor.
Configuring the System

Table of Contents

Other manuals for R&S NRP8S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP8S and is the answer not in the manual?

R&S NRP8S Specifications

General IconGeneral
BrandR&S
ModelNRP8S
CategoryAccessories
LanguageEnglish

Related product manuals