Remote Control Commands
R&S
®
NRPxxS(N)
136User Manual 1177.5079.02 ─ 10
SYSTem:INITialize
Sets the sensor to the standard state, i.e. the default settings for all test parameters
are loaded in the same way as with *RST. The sensor then outputs a complete list of
all supported commands and parameters. With the command, the remote-control soft-
ware can automatically adapt to the features of different types of sensors with different
functionality.
Usage: Event
SYSTem:LANGuage
<language>
Selects an emulation of a different command set.
Parameters:
<language> SCPI
*RST: SCPI
SYSTem:LED:COLor <color>
Sets the color and the flash code of the system status LED, if the operating mode of
the LED is set to USER (SYSTem:LED:MODE).
Parameters:
<color> Hexadecimal code described as
0x0krrggbb
with
k = 0: steady on; k = 1: slowly flashing; k = 2: fast flashing
rr = red
gg = green
bb = blue
In NRP legacy communication, the parameter is a standard deci-
mal number, representing the corresponding hexadecimal code.
Range: 0x00 to 0x02FFFFFF
*RST: 0x00A0A0A0
Example:
SYST:LED:MODE USER
Sets the system status LED operating mode to user.
SYST:LED:COL #H01a00000
The LED flashes slowly in red.
SYSTem:LED:MODE SENSor
Sets the system status LED operating mode back to the sensor
internal settings.
SYSTem:LED:MODE <mode>
Sets whether the color of the system status LED is controlled by the sensor firmware or
by the user settings.
Configuring the System