EasyManuals Logo

R&S NRP8S User Manual

R&S NRP8S
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #137 background imageLoading...
Page #137 background image
Remote Control Commands
R&S
®
NRPxxS(N)
137User Manual 1177.5079.02 ─ 10
For more information, see SYSTem:LED:COLor.
Parameters:
<mode> USER | SENSor
*RST: SENSor
SYSTem:MINPower?
Queries the lower power measurement limit.
This value changes if [SENSe<Sensor>:]CORRection:SPDevice:STATe is set to
ON. The lower measurement limit refers to the sensor or to the combination of a sensor
and the components connected ahead of it.
Use this query to determine a useful resolution for the result display near the lower
measurement limit.
Usage:
Query only
SYSTem:PARameters?
Lists all commands with default values, limits and ranges.
Usage:
Query only
SYSTem:PARameters:DELTa?
Lists all commands that differ from the defined default status set by *RST on page 68.
The commands are output with default values, limits and ranges.
Usage:
Query only
SYSTem:PRESet
Triggers a sensor reset.
The command essentially corresponds to the *RST command, with the exception that
the settings of the following commands are persistently held:
INITiate:CONTinuous
SENSe:AVERage:TCONtrol
SENSe:TRACe:AVERage:TCONtrol
Usage:
Event
SYSTem:REBoot
Reboots the sensor.
Usage:
Event
Configuring the System

Table of Contents

Other manuals for R&S NRP8S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP8S and is the answer not in the manual?

R&S NRP8S Specifications

General IconGeneral
BrandR&S
ModelNRP8S
CategoryAccessories
LanguageEnglish

Related product manuals