Loading...
Remote Control Commands
R&S
®
NRPxxS(N)
70User Manual 1177.5079.02 ─ 10
Example:
*TST?
Query
1
Response: Failed
Usage: Query only
*WAI
WAIt to continue
Prevents the execution of the subsequent commands until all preceding commands
have been executed and all signals have settled.
Usage:
Event
9.4
Preparing for the Measurement
Before starting a measurement, you need to do the following:
Selecting the Reference Source............................................................................. 70
Selecting a Measurement Path...............................................................................70
Selecting a Measurement Mode............................................................................. 71
9.4.1 Selecting the Reference Source
The ROSCillator subsystem contains commands for configuring the reference
source.
[SENSe<Sensor>:]ROSCillator:SOURce <source>
Sets the source of the reference oscillator. Refers typically to a precision, stabilized
time base.
Parameters:
<source> INTernal | EXTernal | HOST
INTernal
Internal precision oscillator
EXTernal | HOST
External signal supplied at the host interface connector.
*RST: INTernal
Example:
ROSC:SOUR INT
9.4.2 Selecting a Measurement Path
The RANGe subsystem contains commands for selection of a measurement path.
Preparing for the Measurement