Remote Control Commands
R&S
®
NRPxxS(N)
69User Manual 1177.5079.02 ─ 10
Stores the current device state under the specified number. The storage numbers 0 to
9 are available.
Setting parameters:
<number> Range: 0 to 9
*RST: 0
Usage: Setting only
*SRE <register>
Service Request Enable
Sets the service request enable register to the specified value. This command deter-
mines under which conditions a service request is triggered.
Parameters:
<register> Range: 0 to 255
*RST: 0
*STB?
STatus Byte query
Returns the contents of the status byte in decimal form.
Usage:
Query only
*TRG
TRiGger
Triggers a measurement. This command is only valid if the power sensor is in the wait-
ing for trigger state and the trigger source is set to BUS
See TRIGger:SOURce BUS.
Usage:
Event
*TST?
Selftest query
Triggers a self test of the instrument and outputs an error code in decimal form. 0 indi-
cates that no errors have occurred.
Example:
*TST?
Query
0
Response: Passed