Page #139 background image
Remote Control Commands
R&S
®
NRPxxS(N)
139User Manual 1177.5079.02 ─ 10
SYSTem:SERRor:LIST[:NEXT]?
Queries the list of all static errors that have occurred but have already been resolved
for the oldest entry and removes it from the queue. The response consists of an error
number and a short description of the error.
Example:
SYST:SERR:LIST?
Query
0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification"
Response
Usage: Query only
SYSTem:SUTime <update_time>
Effective only in the NRP legacy mode.
Sets the status update time. That is the maximum rate in which the power sensor can
output measurement results. Relevant only in continuous measurement mode
(INITiate:CONTinuousON).
Parameters:
<update_time> Range: 0.0 to 10.0
*RST: 10e-3
Default unit: Seconds
SYSTem:TLEVels?
Queries the possible power test levels of the sensor.
Usage:
Query only
SYSTem:TRANsaction:BEGin
Starts a series of settings.
Usage:
Event
SYSTem:TRANsaction:END
Ends a series of settings.
Usage:
Event
SYSTem[:SENSor]:NAME <sensorname>
Sets the name of the sensor according to your requirements. The specified name is
displayed in the web user interface of the network sensors.
Configuring the System

Table of Contents

Other manuals for R&S NRP8S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP8S and is the answer not in the manual?

R&S NRP8S Specifications

General IconGeneral
BrandR&S
ModelNRP8S
CategoryAccessories
LanguageEnglish

Related product manuals