Remote Control Commands
R&S
®
NRPxxS(N)
73User Manual 1177.5079.02 ─ 10
INITiate:ALL
INITiate[:IMMediate]
Starts a single measurement cycle. The sensor changes from the idle state to the wait-
ing for trigger state. As soon as the trigger condition is fulfilled, the sensor begins the
measurement. Depending on the number of trigger events that are required, e.g. for
averaging, the sensor enters the waiting for trigger state several times. Once the entire
measurement is completed, a measurement result is available, and the sensor enters
the idle state again.
Use the command only after the continuous measurement mode has been switched off
(INITiate:CONTinuous OFF).
See also Chapter 9.5, "Controlling the Measurement", on page 72.
Example:
See Chapter 10.3, "Performing a Buffered Continuous Average
Measurement", on page 146.
Usage: Event
INITiate:CONTinuous <state>
Activates/deactivates the continuous measurement mode. In continuous measurement
mode, the power sensor does not go into the idle state after a measurement has been
completed, but immediately executes another measurement cycle.
See also Chapter 9.5.2, "Triggering", on page 73.
Parameters:
<state> ON
Measurements are performed continuously. If a measurement is
completed, the sensor does not return to the idle state but enters
the waiting for trigger state again.
OFF
Ends the continuous measurement mode, and sets the sensor to
the idle state.
*RST: OFF
Example: See Chapter 10.3, "Performing a Buffered Continuous Average
Measurement", on page 146.
Manual operation: See "Measurement" on page 43
9.5.2 Triggering
In a basic continuous measurement, the measurement is started immediately after the
initiate command, see also Chapter 9.5.2.2, "Waiting for a Trigger Event",
on page 74. However, sometimes you want that the measurement starts only if a
specific condition is fulfilled. For example, if a signal level is exceeded, or in certain
time intervals. For these cases, you can define a trigger for the measurement.
Controlling the Measurement

Table of Contents

Other manuals for R&S NRP8S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP8S and is the answer not in the manual?

R&S NRP8S Specifications

General IconGeneral
BrandR&S
ModelNRP8S
CategoryAccessories
LanguageEnglish

Related product manuals