Page #159 background imageLoading...
Page #159 background image
Remote Control Basics
R&S
®
NRPxxS(N)
159User Manual 1177.5079.02 ─ 10
Boolean Parameters
Boolean parameters represent two states. The "ON" state (logically true) is represen-
ted by "ON" or a numeric value 1. The "OFF" state (logically untrue) is represented by
"OFF" or the numeric value 0. The numeric values are provided as the response for a
query.
Example:
Setting command: SENSe:AVERage:COUNt:AUTO ON
Query: SENSe:AVERage:COUNt:AUTO?
Response: 1
Text parameters
Text parameters observe the syntactic rules for mnemonics, i.e. they can be entered
using a short or long form. Like any parameter, they have to be separated from the
header by a white space. For a query, the short form of the text is provided.
Example:
Setting command: TRIGger:SLOPe POSitive
Query: TRIG:SLOP?
Response: POS
Character strings
Enter strings always in quotation marks (' or ").
Example:
Setting command: SENSe:FUNCtion "POWer:AVG"
Query: SENS:FUNC?
Response: "POWer:AVG"
Block data
Block data is a format which is suitable for the transmission of large amounts of data. A
command using a block data parameter has the following structure:
SCPI Command Structure